Machine Solutions Inc.
MSI-eNewsletterSignUp
MSI-Contact
Send this page - Share this Information
Search MSI
Balloon Catheter Pleating and Folding Technology Overview Transcatheter Heart Valve Crimping Tools Overview Balloon Expandable Stent & Polymer Scaffold Crimping Overview Custom Equipment for Medical Device Manufacturing & Testing Overview
Catheter Laminating Technology Overview Mandrel Straightening Equipment Overview Self-Expanding Stent Loading Overview
Catheter Laser Bonding Technology Overview Marker Band Swaging Equipment Overview Thermal Bonding for Medical Device Manufacturing Overview
Medical Device Performance Testing Overview Part Inspection Systems Overview
MSI Home Page What's New at MSI Exhibitions Careers at MSI About MSI Contact MSI

SR1000
Stent Securement Testing Equipment

SR1000 Stent Securement Testing Equipment is designed to measure and record the force required to dislodge a balloon-expandable stent from its proper position on a balloon.
The SR1000 Stent Securement Testing Equipment is designed to measure and record the force required to dislodge a balloon-expandable stent from its proper position on a balloon. The SR1000 is comprised of a multi-sectioned segmental head, including one segment that radially grips the balloon and another segment that captures the stent edge. These head sections separate to dislodge the stent proximally or distally from the balloon. The SR1000 utilizes stepper motor linear actuators for precise diameter control and smooth pull action. Test results can be used for regulatory submissions, competitive product testing, R&D device evaluation and inline manufacturing quality assurance.

MSI designed the stent securement tester to work as a standalone unit with an optional submersible configuration for testing in an aqueous, temperature-controlled environment. The system can also be purchased as an add-on to the MSI Interventional Device Testing Equipment (IDTE2000).

The SR equipment is designed with international testing recommendations in mind. These include: the FDA guidance document titled "Non-Clinical Tests and Recommended Labeling for Intravascular Stents, and Associated Delivery Systems", ASTM F2394-07 Guidance Document for Stent Securement, and ISO Standard 25539-2: 2008.

SR1000 Catalog Page
Testing Equipment Videos
Operation Video
SR1000 Operation Video
Ins and Outs of Stent Securement Testing
Ins & Outs of Stent Securement Testing (20 minutes)
Stent Securement Features:
• Guide-type securement test.
• Proximal or distal displacement of stent.
• High speed data acquisition system.
• Quick set up.
• Repeatable and reproducible data.
• Quick, simple software controlled calibration.
• Graphical displaying of force vs. displacement in real time.
• Diameter control and force feedback on capture segments.
• Video assist with sample alignment & video capture of test.
• Data output setup for spreadsheet analysis.
• Data output includes: profile parameters, peak force and
associated displacement.

Summary of Spec Data:
Capture Head Diameter Range
0.25 - 4.5mm
Capture Head Length
0.5mm
Gripper Head
Diameter Range
0.25 - 4.5mm
Gripper Head Length
6.0mm
Diameter Accuracy
0.02mm
Diameter Resolution
0.01mm
Force Accuracy
0.1N
Force Resolution
0.1N
Pull Rate
0.01– 30.0in/min
Power Requirements
110 or 220 VAC
Capture Head
Load Cell Size
10lb
Gripper Head
Load Cell Size
10lb
Pull Load Cell Size
5lb
Temperature Controlled Water Bath
Optional with
submersible system
General Warranty
1 Year - Routine
Maintenance Required
Stent Securement Base:
• Stainless steel ten segment design.
• Interchangeable handling system for proximal or distal pull.
• Stepper motor driven linear actuators.
• Standard S beam integrated load cells.
 
Control Module Specifications:
• High speed motion control system.
• High speed data acquisition (100 samples/second).
• Ethernet PC to control module connection.
• Custom cable connection between control module and base.
 
Software Specifications:
• Windows PC based software.
• Adjustable testing parameters include: head close speed, capture diameter,
capture maximum force, gripper force, pull speed and pull distance.
• Pull force and distance are measured, displayed, graphed and written to a
CSV (Comma Separated Variable) file.
• Unlimited profile creation and storage.

Product images are representative of equipment offerings and may differ from standard equipment.
© Copyright 2012 All Rights Reserved, Machine Solutions Inc. (MSI)