| Form/Fold/Set
Base Unit Specifications: |
• Precision bushing and guide pin alignment
for all interchangeable
• components (pleat head,
fold head, catheter handling
system)
+/- .003”. |
 |
| • Thermal control adjustable from 25°C
to 150°C. |
 |
| • V-block position sensor activated with optical
solid state
limit/home
switches. |
 |
| • Software controlled open/close diameters
+/- .001”. |
 |
| • Fold force capability 75lbs. |
 |
| • Vacuum to 600 mmHg. |
 |
• Automated servo motor position and advance
catheter handling system,
• hard v-block, 60mm length
with improved clamping
system. |
 |
| • Vision assistance on pleating head (10X). |
 |
| • On/Off key for light curtain. |
 |
| • FasTest connectors with active vacuum indicator
lights. |
 |
| • Fixed height workbench. |
 |
• Electrical transformer to be installed into
e-cabinet for easy conversion to
• European electrical
input (including wiring
diagrams). |
 |
| • Bar Code Scanning Capability. |
 |
| • Light curtain safety shroud. |
 |
| • “V” block modifications with
centering and improved clamping
mechanism. |
| |
| PLC Specifications: |
| • PC controlled. |
 |
| • Touch screen operated with Visual Basic
software. |
 |
| • Pleat/fold head dwell time is programmable
by the user. |
 |
| • Pleat/fold head delay time is programmable
by the user. |
 |
| • Pleat and fold head diameter calibration. |
 |
| • Metric and English units for diameter
and length. |
 |
• Data collection and monitoring of applied
force to activation arms during
• pleating and folding. |
 |
| • Data collecting of diameter settings on
pleat and fold head. |
 |
| • Interlock safety feature monitored by
the PC and halts the machine. |
 |
| • Vacuum delay controlled by PC. |
 |
| • All homing sequences ONLY after power
failure. |
 |
| • Adjustable pleat/fold head close speed
(in. mm/sec)
programmable by user. |
 |
| • Product advance speed (in. mm/sec) programmable
by user. |
 |
| • PC will be internal to e-cabinet. |
| |
| Pleat
Head Specifications: |
• One (1) Wavepleat™ mechanism with
customer specified pleat/segment
• design. |
 |
| • Wavepleat™ mechanism with customer
specified diameters
and length. |
 |
| • One (1) cartridge heater per pleat segment. |
| |
| • LED control box. |
 |
| • Heater elements monitored with PC software. |
| |
| Pleat
Head #2 Specifications: |
• One (1) Wavepleat™ mechanism with
with customer specified pleat/segment
• design. |
 |
| • Wavepleat™ mechanism customer specified
diameters and length. |
 |
| • One (1) cartridge heater per pleat segment. |
 |
| • LED control box. |
 |
| • Heater elements monitored with PC software. |
| |
| Fold
Head Specifications: |
| • One (1) Folding mechanism with open diameter
of 6.0 mm,
60 mm length. |
 |
| • Ten-segment stainless steel fold head
design. |
 |
| • Ten (10) cartridge heaters per fold head. |
 |
| • Heater elements monitored with PC software. |
 |
| • Force monitoring from activation load
cells. |